Assessment of the Diagnostic Performance of Fully Automated Hepatitis E Virus (HEV) Antibody Tests.

Anna EichhornFranziska NeumannCarina BäumlerImke GutsmannOlaf GrobeFrieda SchlüterSina MüllerAndi Krumbholz
Published in: Diagnostics (Basel, Switzerland) (2024)
The detection of anti-hepatitis E virus (HEV) antibodies contributes to the diagnosis of hepatitis E. The diagnostic suitability of two automated chemiluminescence immunoassays (CLIAs, LIAISON ® MUREX Anti-HEV IgG/Anti-HEV IgM test, DiaSorin) was assessed by comparison with the results of a combination of enzyme immunoassays and immunoblots ( recom Well HEV IgG/IgM ELISA, recom Line HEV IgG/IgM, MIKROGEN). Samples with a deviating result were analyzed with the WANTAI ELISAs. Compared to the recom Well ELISAs, the Anti-HEV IgG CLIA had a percentage overall agreement (POA) of 100% (149/149; 95% CI: 97.5-100%) and the Anti-HEV IgM CLIA had a POA of 83.3% (85/102; 95% CI: 74.9-89.3%); considering the recom Line HEV IgM results, the POA was 71.7% (38/53; 95% CI: 58.4-82%). The WANTAI test confirmed 52.9% (9/17) of negative CLIA IgMs; HEV RNA was not detectable. Since acute infection with the Epstein-Barr virus (EBV) or human cytomegalovirus (CMV) may influence the results of other serological assays, HEV antibodies were examined in 17 EBV and 2 CMV patients: One had an isolated and probably unspecific HEV IgM in the CLIA, as HEV RNA was not detectable. Both CLIAs are well suited for HEV diagnostics, but isolated IgM should be confirmed. An acute EBV/CMV infection can influence HEV serodiagnostics.