Performance of the BD-FACS Presto for CD4 count and hemoglobin measurement in a district hospital and rural laboratory in Ghana.

Zelda R MoranJilian A SacksFrancis Kwabena FrimpongAtta Boahen FrimpongYanis Ben Amor
Published in: PloS one (2019)
Overall, the Presto gave higher estimates of CD4 count compared to FACSCount, and hemoglobin measurements were higher than from Sysmex. Samples of capillary blood in turn gave higher results for both measurements compared to venous blood, consistent with previous analyses. These findings should be considered when selecting CD4 count machines for use at the point of care, especially in remote areas where capillary blood sampling may be preferable, but are likely balanced by device's ease of use, portability, and ability to expand access to services. These results are some of the first to demonstrate the accuracy of the FACSPresto in West Africa and show that this device can be successfully operated in a very rural lab setting and may therefore assist to provide CD4 count and hemoglobin concentration measurement to populations in need.