The word-with-noise test: development, validation and reference values.
Lidiéli Dalla CostaAna Valéria de Almeida VaucherKarina Carlesso PagliarinMaristela Julio CostaPublished in: CoDAS (2024)
The Word-with-Noise Test consists of lists of monosyllabic and disyllabic words and speech spectrum noise (Stage 1). The application strategy for the test was defined as the determination of the Speech Recognition Threshold with a fixed noise level at 55 dBHL (Stage 2). Regarding criterion validity, the instrument demonstrated adequate ability to distinguish between normal-hearing subjects and subjects with hearing loss (Stage 3). Reference values for the test were established as cut-off points expressed in terms of signal-to-noise ratio: 1.47 dB for the monosyllabic stimulus and -2.02 dB for the disyllabic stimulus. Conclusion: The Word-with-Noise Test proved to be quick to administer and interpret, making it a useful tool in audiological clinical practice. Furthermore, it showed satisfactory evidence of criterion validity, with established reference values.