Simultaneous quantification of imatinib and its main metabolite N-demethyl-imatinib in human plasma by liquid chromatography-tandem mass spectrometry and its application to therapeutic drug monitoring in patients with gastrointestinal stromal tumor.

Wei ZhuangHai-Bo QiuXin-Meng ChenXiu-Hong YuanLi-Fang YangXiao-Wei SunXiao-Jun ZhouMin HuangXue-Ding WangZhi-Wei Zhou
Published in: Biomedical chromatography : BMC (2017)
The aim of this study was to improve and validate a more stable and less time-consuming method based on liquid chromatography and tandem mass spectrometry (LC- MS/MS) for the quantitative measurement of imatinib and its metabolite N-demethyl-imatinib (NDI) in human plasma. Separation of analytes was performed on a Waters XTerra RP18 column (50 × 2.1 mm i.d., 3.5 μm) with a mobile phase consisting of methanol-acetonitrile-water (65:20:15, v/v/v) with 0.05% formic acid at a flow-rate of 0.2 mL/min. The Quattro MicroTM triple quadruple mass spectrometer was operated in the multiple-reaction-monitoring mode via positive electrospray ionization interface using the transitions m/z 494.0 → 394.0 for imatinib, m/z 479.6 → 394.0 for NDI and m/z 488.2 → 394.0 for IS. The method was linear over 0.01-10 μg/mL for imatinib and NDI. The intra- and inter-day precisions were all <15% in terms of relative standard deviation, and the accuracy was within ±15% in terms of relative error for both imatinib and NDI. The lower limit of quantification was identifiable and reproducible at 10 ng/mL. The method was sensitive, specific and less time-consuming and it was successfully applied in gastrointestinal stromal tumor patients treated with imatinib.