Development of electrostatic-induced charge detector for multiturn time-of-flight mass spectrometer.

Ken-Ichi BajoJun AokiMorio IshiharaShizuho FuruyaMasahiro NishimuraMiwa YoshitakeHisayoshi Yurimoto
Published in: Journal of mass spectrometry : JMS (2022)
We developed an autocorrelation function to resolve the overtaking problem in a multiturn time-of-flight mass spectrometer (TOF-MS). The function analyzes the characteristic period for one lap of each ion packet and derives a mass spectrum from a signal pulse train composed of multiturn ion packets. To detect the ion pulse train, a new nondestructive ion detector was developed and installed in the multiturn orbit of MULTUM-S II. This detector is composed of an electrostatically induced charge detector, a preamplifier, and a digitizer. The electrostatic noises are smaller than the single-ion signals owing to the accumulation of the multiturn TOF spectrum. The conventional ion detector of TOF-MS is operated after collecting the signal pulse train. The multiturn TOF spectrum was convolved with an autocorrelation function to derive the mass spectrum. The convolved mass spectrum performed a mass resolving power (MRP) of 28,200 at m/z 69 and mass accuracy of 28 ppm for the perfluorotributylamine (PFTBA) gas sample.