Login / Signup

Quasistatic Solutions versus Full-Wave Solutions of Single-Channel Circular RF Receive Coils on Phantoms of Varying Conductivities at 3 Tesla.

Michael J BeckDennis L ParkerJ Rock Hadley
Published in: Concepts in magnetic resonance. Part B, Magnetic resonance engineering (2021)
Full-wave simulations were on average 3x more accurate than the quasistatic simulations. Full-wave simulations were more accurate in characterizing the wave effects within the sample, though they were not able to fully account for the skin effect when calculating coil noise.
Keyphrases
  • molecular dynamics
  • monte carlo
  • high resolution
  • magnetic resonance
  • air pollution
  • mass spectrometry
  • wound healing