Analysis of opacification patterns in intraocular lenses (IOL).
Marc MackertDaniel Rudolf MuthEfstathios VounotrypidisConstanze DegerDavid GoldblumMehdi ShajariPascal Willy HaslerSiegfried PriglingerArmin WolfPublished in: BMJ open ophthalmology (2021)
In our series, the morphology of IOL opacification did not follow the existing pathogenetic classification that strictly discriminates between primary and secondary causes. Fine-granular IOL opacification occurs with similar patterns in both type 1 and type 2 IOL opacification, while a crust-like pattern was only detected in type 2 IOL opacifications. Consequently, susceptibility of an IOL to opacification is caused by a multifactorial combination of material and processing properties as well as individual (pathological) conditions of the patient.