Login / Signup

n-Butylamine for Improving the Efficiency of Untargeted Mass Spectrometry Analysis of Plasma Metabolite Composition.

Dmitry L MaslovOxana P TrifonovaElena E BalashovaPetr G Lokhov
Published in: International journal of molecular sciences (2019)
A comparative study of the impact of n-butylamine and traditionally used additives (ammonium hydroxide and formic acid) on the efficiency of the electrospray ionization (ESI) process for the enhancement of metabolite coverage was performed by direct injection mass spectrometry (MS) analysis in negative mode. Evaluation of obtained MS data showed that n-butylamine is one of the most effective additives for the analysis of metabolite composition in ESI in negative ion mode (ESI(-)) The limitations of the use of n-butylamine and other alkylamines in the analysis of metabolic composition and a decontamination procedure that can reduce MS device contamination after their application are discussed. The proposed procedure allows the performance of high-sensitivity analysis of low-molecular-weight compounds on the same MS device in both polarities.
Keyphrases