Evaluation of dry textile electrodes for long-term electrocardiographic monitoring.

Milad Alizadeh MeghraziBinbin YingAlessandra SchlumsEmily LamLadan EskandarianFarhana AbbasGurjant SidhuAmin MahnamBastien MoineauMilos R Popovic
Published in: Biomedical engineering online (2021)
Dry textile electrodes demonstrated similar impedance when in raised or flat structures. The tested pressure variations had an insignificant impact on electrode impedance. Looking at the effect of impedance on ECG signals, a noticeable effect on ECG signal performance metrics was not observed. Therefore, it is suggested that impedance alone is possibly not the primary indicator of signal quality. As well, the developed methods can also serve as useful guidelines for future textile dry-electrode design and testing for practical ECG monitoring applications.