Single-energy pediatric chest computed tomography with spectral filtration at 100 kVp: effects on radiation parameters and image quality.
Boris BodelleConstanze FischbachChristian BoozIbrahim YelClaudia FrellesenMoritz KaupMartin BeeresThomas J VoglJan-Erik ScholtzPublished in: Pediatric radiology (2017)
Using 100-kVp spectral filtration chest CT in children by means of a tube-based tin-filter on a third-generation dual-source CT scanner increases effective tube current up to a factor of 10 to provide similar image quality at equivalent dose compared to standard single-source CT without spectral filtration.