Login / Signup

Coupled transfer function model for the evaluation of implanted cables safety in MRI.

Julie KabilJacques FelblingerPierre-André VuissozAlexia Missoffe
Published in: Magnetic resonance in medicine (2020)
It is necessary to take into account the coupling effect between leads to evaluate the safety of implanted devices. This work shows that, in the case of 2 cables separated by a constant distance, that an experimentally determined coupled transfer function allows estimation of the heating of the 2 electrodes for a given incident field. Further work should take into account the in vivo varying distance between the 2 cables.