Pelvic Ring Fractures: A Biomechanical Comparison of Sacral and Lumbopelvic Fixation Techniques.
Sudharshan TripathiNorihiro NishidaSophia SoehnlenAmey KelkarYogesh KumaranToshihiro SekiTakashi SakaiVijay K GoelPublished in: Bioengineering (Basel, Switzerland) (2024)
Longer sacral screw fixations were superior in stabilizing sacral and contralateral sacroiliac joint range of motion. Lumbopelvic fixations displayed a higher degree of stabilization in the horizontal displacement compared to vertical displacement of pubic rami fracture, while also indicating the highest rod stresses. When determining the surgical approach for pelvic ring fractures, patient-specific factors should be accounted for to weigh the advantages and disadvantages for each technique.