Login / Signup

Achievable aspiration flow rates with large balloon guide catheters during carotid artery stenting.

Tilman SchubertLeonardo Rivera-RiveraAlejandro Roldan-AlzateDaniel ConsignyLorenz LeitnerCharles StrotherBeverly Aagaard-Kienitz
Published in: CVIR endovascular (2020)
Based on the results of this study, flow reversal in the ICA during common carotid artery occlusion is most likely achieved with the smallest diameter stent sheath and the stent model with the shortest outer stent sheath maximum diameter. This implies that embolic protection during emergency CAS through aspiration is most effective with these models.
Keyphrases
  • ultrasound guided
  • optic nerve
  • emergency department
  • public health
  • crispr cas
  • healthcare
  • antiplatelet therapy
  • coronary artery disease
  • acute coronary syndrome
  • atrial fibrillation