Impacts of acquisition and reconstruction parameters on the absolute technetium quantification of the cadmium-zinc-telluride-based SPECT/CT system: a phantom study.
Ruyi ZhangMiao WangYaqian ZhouShen WangYiming ShenNing LiPeng WangJian TanZhaowei MengQiang JiaPublished in: EJNMMI physics (2021)
CZT-SPECT/CT of technetium resulted in good quantification accuracy. The favourable acquisition parameters might be a 15% energy window and 40 s/frame of acquisition time. The favourable reconstruction parameters might be 35 iterations, 20 subsets, the AC + SC + RR correction combination, and no filter.