Flow irregularities from syringe infusion pumps caused by syringe stiction.
Neris PeduzziMartin BatlinerBeate GrassPhilipp Karl BuehlerMarianne Schmid DanersMirko MeboldtMarkus WeissPublished in: Paediatric anaesthesia (2020)
Infusion set rate rather than syringe brand or filling volume was a major predictor for syringe stiction-related amount of variation in the calculated plasma epinephrine concentration.