Prospective Comparison of Transient Elastography Using Two Different Devices: Performance of FibroScan and FibroTouch.
Joao Tiago SerraJohannes MuellerHaidong TengOmar El ShaarawySebastian MuellerPublished in: Hepatic medicine : evidence and research (2020)
Although LS by FT shows good correlation with LS-FS, it has larger variation, continuously overestimates LS and completely fails in ascites. Moreover, FT-UAP seems to be a misleading parameter for steatosis assessment because it is at least in part calculated from mandatory entered patient data. In conclusion, novel LS cut-off values need to be defined for LS-FT and usage of UAP is not recommended.