Modes of administration of nitric oxide devices and ventilators flow-by impact the delivery of pre-determined concentrations.

Alice VuillermozMathilde LefrancNathan ProuvezClément BraultYoann ZerbibMary SchmittJean-Marie ForelMathieu Le TutourArnaud LesimpleAlain MercatJean-Christophe RichardFrançois M Beloncle
Published in: Annals of intensive care (2024)
We identified three different generations of iNO-devices, based on their gas administration modalities, that were associated with highly variable iNO concentrations' accuracy. Ventilator's flow by significantly impacted iNO concentration. Only the Proportional devices permitted to accurately deliver iNO whatever the conditions and the ventilators tested.