Three-dimensional optimization and sensitivity analysis of dental implant thread parameters using finite element analysis.

Maryam GeramizadehHamidreza KatoozianReza AmidMahdi Kadkhodazadeh
Published in: Journal of the Korean Association of Oral and Maxillofacial Surgeons (2018)
Based on the results of this study, the optimal implant design has micro-threads with 0.307 and 0.286 mm depth and pitch, respectively, in the upper area and V-shaped threads with 0.405 and 0.808 mm depth and pitch in the rest of the body. These results indicate that micro-thread parameters have a greater effect on stress and strain values.
Keyphrases
  • finite element analysis
  • optical coherence tomography
  • soft tissue
  • image quality
  • magnetic resonance imaging
  • computed tomography
  • stress induced