Login / Signup

Time-Domain Investigation of Switchable Filter Wide-Band Antenna for Microwave Breast Imaging.

Amir HaiderMuhibUr RahmanMahdi NaghshvarianJahromiHyung Seok Kim
Published in: Sensors (Basel, Switzerland) (2020)
This paper investigates the time-domain performance of a switchable filter impulse radio ultra-wideband (IR-UWB) antenna for microwave breast imaging applications. A miniaturized CPW-fed integrated filter antenna with switchable performance in the range of the Worldwide Interoperability for Microwave Access (WiMAX) and Wireless Local Area Network (WLAN) bands could operate well within a 3.0 to 11 GHz frequency range. The time-domain performance of the filter antenna was investigated in comparison to that of the designed reference wideband antenna. By comparing both antennas' time-domain characteristics, it was seen that the switchable filter antenna had good time-domain resolution along with the frequency-domain operation. Additionally, the time-domain investigation revealed that the switchable filter wide-band antenna performed similarly to the reference wide band antenna. This antenna was also utilized for a tumor detection application, and it was seen that the switchable filter wide-band antenna could detect a miniaturized irregularly shaped tumor easily, which is quite promising. Such an antenna with a good time-domain resolution and tumor detection capability will be a good candidate and will find potential applications in microwave breast imaging.
Keyphrases
  • energy transfer
  • high resolution
  • single molecule
  • climate change
  • quantum dots
  • loop mediated isothermal amplification
  • fluorescence imaging
  • sensitive detection