Safety and feasibility of a factory-calibrated continuous glucose monitoring system in term and near-term infants at risk of hypoglycemia.
Eri NishimuraShuntaro OkaJunichi OzawaKosuke TanakaTaichi MomoseKazuhiko KabeFumihiko NambaPublished in: Turkish archives of pediatrics (2021)
The factory-calibrated CGMS was a safe and feasible modality for glucose monitoring. However, it has a tendency to overestimate the blood glucose concentrations. Therefore, this system should be used cautiously for neonates at risk of hypoglycemia, especially within 3 hours after sensor placement.