Data Quality and Optimal Background Correction Order of Respiratory-Gated k-Space Segmented Spoiled Gradient Echo (SGRE) and Echo Planar Imaging (EPI)-Based 4D Flow MRI.

Federica ViolaPetter DyverfeldtCarl-Johan CarlhällTino Ebbers
Published in: Journal of magnetic resonance imaging : JMRI (2019)
Both sequences yielded good internal data consistency when an adequate background correction was applied. Second and first BC order were considered sufficient for transvalvular flow analysis in SGRE and EPI, respectively. Higher BC orders were preferred for particle tracing. Level of Evidence 4 Technical Efficacy Stage 1 J. Magn. Reson. Imaging 2020;51:885-896.