Comparison of IOP Measurement by Goldmann Applanation Tonometer, ICare Rebound Tonometer, and Tono-Pen in Keratoconus Patients after MyoRing Implantation.

Mahmoud F RatebMahmoud Abdel-RadiZeiad EldalyMohamed Nagy ElmohamadyAsaad Noor El Din
Published in: Journal of ophthalmology (2019)
Seventeen eyes (10 patients) were evaluated. The mean IOP reading was 13.9 ± 3.68, 12.41 ± 2.87, and 14.29 ± 1.31 mmHg in GAT, ICare, and Tono-Pen group, respectively. There was a significant difference between IOP readings by GAT/ICare and Tono-Pen/ICare (p value: 0.032 and 0.002, respectively) with no significant difference between GAT/Tono-Pen (p value: 0.554). Mean difference in IOP measurements between GAT/ICare was 1.49 ± 2.61 mmHg, Tono-Pen/ICare was 1.89 ± 2.15 mmHg, and GAT/Tono-Pen was -0.39 ± 2.59 mmHg. There was no significant correlation between the difference in IOP readings among any pair of devices and CCC or TCL. The Bland-Altman analysis showed a reasonable agreement between any pair of tonometers.