Is load control necessary to produce physiological AP displacement and axial rotation in wear testing of TAR?
Nathan C HoColin P McCartySang-Hyun ParkJoan R WilliamsNeda Farbod GilmartinEdward EbramzadehSophia N SangiorgioPublished in: Journal of orthopaedic research : official publication of the Orthopaedic Research Society (2020)
The International Standard Organization, ISO 22622, specifies two options for joint wear simulator evaluation of total ankle replacements (TARs): load-controlled and displacement-controlled. In the present study, the load-controlled testing parameters were applied to cadaveric specimens to quantify and compare the observed sagittal translations and axial rotations to those specified under the displacement-controlled option. Twelve cadaveric specimens were stripped of extraneous tissues, keeping surrounding ankle ligaments. A halo was used to produce plantarflexion and dorsiflexion of the talus through two screws, while a baseplate resisted axial loads. The axial force and torque were applied to the tibia and fibula under force and torque feedback control. An anterior-posterior force was applied to the tibia. Plantarflexion-dorsiflexion were applied using rotation control. To protect the cadaveric specimens, loads were applied at 50% of the specified load profile while plantarflexion-dorsiflexion rotation was applied as specified. There was variation among specimens in magnitudes of anterior-posterior displacement with peaks ranging from 3.3 mm posteriorly to 3.0 mm anteriorly. Likewise, there was variation among specimens in magnitude of axial rotation, with peaks ranging from 11° external rotation to 4.5° internal rotation. However, the mean magnitudes of AP displacement and axial rotation did not exceed those specified by ISO 22622.