Causes of impaired biventricular pacing in cardiac resynchronization devices with left ventricular sensing.
Andreas HaeberlinSylvain PlouxAntoine NoelRémi ChauvelNicolas WelteHugo MarchandMichel HaïssaguerrePhilippe RitterRomain EschalierPierre BordacharPublished in: Pacing and clinical electrophysiology : PACE (2020)
Inadequate programming (active LV sensing with T-wave protection) is the main cause of impaired resynchronization in devices with LV sensing. In general, we recommend the deactivation of the LV T-wave protection function.