Login / Signup

Highly Sensitive, Portable Detection System for Multiplex Chemiluminescence Analysis.

Yannan YuWei NieKaiqin ChuXi WeiZachary J Smith
Published in: Analytical chemistry (2023)
Chemiluminescence (CL) has emerged as a critical tool for the sensing and quantification of various bioanalytes in virtually all clinical fields. However, the rapid nature of many CL reactions raises challenges for typical low-cost optical sensors such as cameras to achieve accurate and sensitive detection. Meanwhile, classic sensors such as photomultiplier tubes are highly sensitive but lack spatial multiplexing capabilities and are generally not suited for point-of-care applications outside a standard laboratory setting. To address this issue, in this paper, a miniaturized and versatile silicon-photomultiplier-based fiber-integrated CL device (SFCD) was designed for sensitive multiplex CL detection. The SFCD comprises a silicon photomultiplier array coupled to an array of high numerical aperture plastic optical fibers to achieve 16-plex detection. The optical fibers ensure efficient light collection while allowing the fixed detector to be mated with diverse sample geometries (e.g., circular or grid), simply by adjusting the fiber configuration. In a head-to-head comparison with a lens-based camera system featuring a cooled detector, the SFCD achieved a 14-fold improved limit of detection in both direct and enzyme-mediated CL reactions. The SFCD also features improved compactness and lower cost, as well as faster temporal resolution compared with camera-based systems while preserving spatial multiplexing and good environmental robustness. Thus, the SFCD has excellent potential for point-of-care biosensing applications.
Keyphrases