High rate episodes in a dual chamber pacemaker: What is the mechanism?

Vijay Shekar PVickram Vignesh RangaswamyMuthiah SubramaniamAuras R AtreyaSachin YalagudriSridevi ChennapragadaDaljeet Kaur SagguVickram V Rangaswamy
Published in: Pacing and clinical electrophysiology : PACE (2022)
Detection of high rate episodes can potentially result from oversensing of physiological or nonphysiological signals. Assessment of lead integrity, lead connection and analysis of noise characteristics on electrograms are decisive steps in the evaluation of oversensing. We report a case of high atrial and ventricular rate episodes due to minute ventilation oversensing in the presence of lead connector issues.