Temperature dependence, accuracy, and repeatability of T1 and T2 relaxation times for the ISMRM/NIST system phantom measured using MR fingerprinting.

Ben K StattonJoely SmithMary E FinneganGregor KörzdörferRebecca A QuestMatthew Grech-Sollars
Published in: Magnetic resonance in medicine (2021)
MRF demonstrated relaxation times with a temperature dependence similar to that of conventional mapping methods. Temperature-corrected T1 and T2 values from both dictionaries showed adequate accuracy and excellent repeatability in this phantom study.