Login / Signup

An evaluation of a coagulation system (Xprecia Stride) for utilisation in anticoagulation management.

Deborah McCahonAndrea RoalfeDavid A Fitzmaurice
Published in: Journal of clinical pathology (2017)
INRs tested using the Xprecia Stride system showed good agreement with the laboratory and CoaguChek systems. Findings indicate that in the hands of the intended users the Xprecia Stride is accurate, reliable and acceptable for use in a routine clinical setting.
Keyphrases
  • atrial fibrillation
  • venous thromboembolism
  • clinical practice
  • high resolution