Commissioning of optical surface imaging systems for cranial frameless stereotactic radiosurgery.
Lei ZhangSarath VijayanSheng HuangYulin SongTianfang LiXiang LiElizabeth HippMaria F ChanHsiang-Chi KuoXiaoli TangGrace TangSeng Boh LimDale Michael LovelockAse BallangrudGuang LiPublished in: Journal of applied clinical medical physics (2021)
A commissioning method is recommended using the 3D plate calibration, which is verified by radiation isocenter and validated with couch-angle dependency, baseline drift, and gated-delivered-dose equivalency tests. This method characterizes OSI uncertainties, ensuring motion-monitoring accuracy for SRS treatments.