Impact of super-resolution and image acquisition on the detection of calcifications in digital breast tomosynthesis.
Bruno BarufaldiRaymond J AcciavattiEmily F ConantAndrew D A MaidmentPublished in: European radiology (2023)
• Super-resolution improves calcification detection and SNR in tomosynthesis; specifically, with the use of smaller reconstruction voxels. • Calcification detection using step-and-shoot motion is superior to that using continuous tube motion. • A detector element size of 70 µm does not provide better detection than 140 µm for small calcifications at the threshold of detectability.