Evidence-Based Consensus Guidelines Series for MicroPulse Transscleral Laser Therapy: Dosimetry and Patient Selection.

Tomas M GrippoRonald M P C de CromMichael GiovingoMarc Töteberg-HarmsBrian A FrancisBrian JerkinsJacob W BrubakerNathan M RadcliffeJella A AnRobert Noecker
Published in: Clinical ophthalmology (Auckland, N.Z.) (2022)
Micropulse TLT is a useful addition to the glaucoma armamentarium. When used with proper surgical technique at energy settings within the boundaries described in this manuscript, MicroPulse TLT is a safe and effective treatment for many types and stages of glaucoma. Based on current knowledge and experience, the consensus recommendation of this expert panel is that the standard MicroPulse TLT settings using the revised MicroPulse P3 Probe should be 2500 mW, 31.3% duty cycle, and 4 sweeps at a sweep velocity of 20 seconds each per hemisphere. Both hemispheres avoiding the 3 and 9 clock hours should be treated. The panel also reached consensus on patient selection for MicroPulse TLT providing guidance for the use of the procedure.