0° vs. 180° CT localiser: The effect of vertical off-centring, phantom positioning and tube voltage on dose optimisation in multidetector computed tomography.

Yazan Al-HayekXiaoming ZhengRob DavidsonChristopher M HayreDana S Al-MousaCampbell FinlayKelly Maree Spuur
Published in: Journal of medical radiation sciences (2021)
Phantom off-centring and localiser direction evidenced large dose variation. It is recommended that the 0° localiser is employed during CT examinations, in order to minimise the potential additional radiation dose which may result from off-centring and the use of lower tube voltages where clinically appropriate.