Login / Signup

The Effect of Paper on the Detection Limit of Paper-Based Potentiometric Chloride Sensors.

Eliza J HerreroBlair K TroudtPhilippe Bühlmann
Published in: Analytical chemistry (2022)
While paper is an excellent material for use in many other portable sensors, potentiometric paper-based sensors have been reported to perform worse than conventional rod-shaped electrodes, in particular in view of limits of detection (LODs). Reported here is an in-depth study of the lower LOD for Cl - measurements with paper-based devices comprising AgCl/Ag transducers. Contamination by Cl - from two commonly used device materials─a AgCl/Ag ink and so-called ashless filter paper─was found to increase the concentration of Cl - in paper-contained samples far above what is expected for the spontaneous dissolution of the transducer's AgCl, thereby worsening lower LODs. In addition, for the case of Ag + , the commonly hypothesized adsorption of metal cations onto filter paper was found not to significantly affect the performance of AgCl/Ag transducers. We note that in the context of chemical analysis, metal impurities of paper are often mentioned in the literature, but Cl - contamination of paper has been overlooked.