CoaguChek® XS versus standard laboratory prothrombin time for anticoagulant monitoring in patients with antiphospholipid syndrome.
Maria Ester S FonsecaGustavo Guimarães Moreira BalbiFlavio SignorelliChristiane P GouveaDanieli Castro de Oliveira AndradePublished in: Lupus (2022)
Although CoaguChek XS and venous laboratory INR demonstrated a good linear correlation in the group of INR ≤4, extra caution should be taken in APS patients, since a reasonable proportion of patients can present differences in INR results that are not acceptable. We do not recommend routine POC in APS patients.