Diagnostic performance of the Sofia® influenza A+B fluorescent immunoassay in adult outpatients in Northern Taiwan.

Jeng-How YangPo-Yen HuangShian-Sen ShieShuan YangKuo-Chien TsaoTsu-Lan WuTing-Shu WuChing-Tai Huang
Published in: Journal of medical virology (2018)
To evaluate the diagnostic performance of the Sofia influenza A+B fluorescent immunoassay (Sofia FIA), we performed a prospective study at the Chang Gung Memorial Hospital in Taiwan from January 2012 to December 2013. Patients who presented at out-patient clinics or the emergency department with influenza-like illness were included. Upper respiratory tract specimens were collected from oropharynx or nasopharynx. Performance of the Sofia FIA was compared to that of the Formosa One Sure Flu A/B Rapid Test. A Real-time reverse transcriptase-polymerase chain reaction assay (RT-PCR) and/or virus culture were used as reference standards. Of the 109 enrolled patients, the sensitivity, specificity, positive, and negative predictive values of the Sofia FIA to detect influenza A virus were 82%, 89%, 77%, and 89%, respectively. These parameters were 100% when the samples were from nasopharynx. The positive predictive value for influenza B virus detection was 29%. The sensitivity of the Sofia FIA for detection of influenza A virus was 93% between days 2 and 4 after onset of symptoms. For specimens with low viral loads (RT-PCR cycle threshold between 30 and 34.9), the sensitivity of The Sofia FIA was 83% (10/12). The Sofia FIA performed effectively in detecting influenza A virus infection. With nasopharyngeal samples, the performance was comparable to RT-PCR. Although influenza viral load typically decreases with time, the Sofia FIA was sensitive enough to identify influenza infecting patients presenting after several days of illness. However, a high false positive rate limits the assay's usefulness to identify influenza B virus infection.
Keyphrases