Clinical validation and evaluation of a novel six-lead handheld electrocardiogram recorder compared to the 12-lead electrocardiogram in unselected cardiology patients (EVALECG Cardio).
Mohammad AzramNoura AhmedLucy LeeseMatthew BrighamRobert BowesStephen B WheatcroftMarcus NgantchaBerthold StegemannGeorge CrowtherMuzahir H TayebjeePublished in: European heart journal. Digital health (2021)
Several parameters recorded by the Kardia 6L (QT interval in all six leads, rhythm detection, PR interval, QRS duration, and cardiac axis) perform closely to the gold standard 12-lead ECG. However, that consistency weakens for left ventricular hypertrophy, QRS amplitudes (Lead I and AVL), and ischaemic changes.
Keyphrases
- left ventricular
- end stage renal disease
- cardiac resynchronization therapy
- ejection fraction
- newly diagnosed
- heart rate
- heart failure
- chronic kidney disease
- peritoneal dialysis
- atrial fibrillation
- acute myocardial infarction
- hypertrophic cardiomyopathy
- aortic stenosis
- heart rate variability
- blood pressure
- percutaneous coronary intervention
- patient reported outcomes
- transcatheter aortic valve replacement
- aortic valve
- label free