The effects of field strength on stimulated echo and motion-compensated spin-echo diffusion tensor cardiovascular magnetic resonance sequences.
Andrew D ScottKe WenYaqing LuoJiahao HuangSimon GoverRajkumar SoundarajanPedro F FerreiraDudley J PennellSonia Nielles-VallespinPublished in: Journal of cardiovascular magnetic resonance : official journal of the Society for Cardiovascular Magnetic Resonance (2024)
While 3T provides benefits in terms of SNR, both STEAM and MCSE can be performed at 1.5T. However, MCSE is unreliable in diastole at both field strengths and STEAM benefits from the improved SNR at 3T over 1.5T. Future clinical research studies may be able to leverage the wider availability of 1.5T CMR hardware where MCSE acquisitions are desirable.