Aptamer-Based Solution-Gated Graphene Transistors for Highly Sensitive and Real-Time Detection of Thrombin Molecules.

Haiyang YuZheng ZhaoBichen XiaoMinghua DengZhaoliang WangZiqin LiHuibin ZhangLei ZhangJingwen QianJinhua Li
Published in: Analytical chemistry (2021)
Thrombin is an important biomarker for various diseases and biochemical reactions. Rapid and real-time detection of thrombin that quickly neutralizes in early coagulation in the body has gained significant attention for its practical applications. Solution-gated graphene transistors (SGGTs) have been widely studied due to their higher sensitivity and low-cost fabrication for chemical and biological sensing applications. In this paper, the ssDNA aptamer with 29 bases was immobilized on the surface of the gate electrode to specifically recognize thrombin. The SGGT sensor achieved high sensitivity with a limit of detection (LOD) up to fM. The LOD was attributed to the amplification function of SGGTs and the suitable aptamer choice. The ssDNA configuration folding induced by thrombin molecules and the electropositivity of thrombin molecules could arouse the same electrical response of SGGTs, helping the device obtain a high sensitivity. The channel current variation of sensors had a good linear relationship with the logarithm of thrombin concentration in the range of 1 fM to 10 nM. The fabricated device also demonstrated a short response time to thrombin molecules, and the response time to the 1 fM thrombin molecules was about 150 s. In summary, the sensing strategy of aptamer-based SGGTs with high sensitivity and high selectivity has a good prospect in medical diagnosis.
Keyphrases