Login / Signup

A New Approach to Accuracy Evaluation of Single-Tooth Abutment Using Two-Dimensional Analysis in Two Intraoral Scanners.

Jiyoun MaengYoung-Jun LimBong-Ju KimMyung-Joo KimHo-Beom Kwon
Published in: International journal of environmental research and public health (2019)
The aim of this study was to two-dimensionally evaluate deviation errors at five digital cross-sections of single-tooth abutment in regards to data obtained from two intraoral scanners, and to evaluate accuracy of individual scanners. Two intraoral scanners, the Trios 3® (3 Shape, Copenhagen, Denmark) and EzScan® (Vatech, Hwaseong, Korea), were evaluated by utilizing 13 stone models. The superimposed 3D data files were sectioned into five different planes: buccal-lingual section (BL), mesial-distal section (MD), transverse high section (TH), transverse middle section (TM), and transverse low section (TL). Accuracy comparison between the two scanners in 5 groups was performed. BL vs. MD of each scanner, and three transverse groups (TH, TM, TL) of each scanner were analyzed for accuracy comparison. In comparison of 2-D analyses for two intraoral scanners, Trios 3® showed statistically significant higher accuracy in root mean square (RMS) at BL, TH, and TL (p < 0.05). For each scanner, RMS value showed that mesial-distal sections were more prone to error than buccal-lingual section, which exhibited statistically significant errors (p < 0.05) while the transverse groups did not. Two-dimensional analysis is more insightful than three-dimensional analysis on single-tooth abutment. In mesiodistal areas, rough prepped areas, and sharp edges where scanner accessibility is difficult, high deviation errors are shown.
Keyphrases