Login / Signup

Inappropriate shocks due to air entrapment in patients with subcutaneous implantable cardioverter-defibrillator: A meta-summary of case reports.

Michele IavaroneAnna RagoGerardo NigroPaolo GolinoVincenzo Russo
Published in: Pacing and clinical electrophysiology : PACE (2022)
IAS due to air entrapment is an early complication of S-ICD implant. No association was observed between the implant technique and air retention. Diagnosis is confirmed by chest x-ray and device interrogation. Management includes switching off the device or, preferably, changing the sensing vector.
Keyphrases
  • soft tissue
  • case report
  • magnetic resonance