Diagnostic accuracy of a high-sensitivity troponin I assay and external validation of 0/3 h rule out strategies.
Patricia van den BergPaul O CollinsonNiall MorrisJessica JanePublished in: European heart journal. Acute cardiovascular care (2022)
With serial sampling over 3 h, the Siemens TNIH assay should be used with a validated algorithm incorporating bespoke cut-offs and absolute delta criteria. In our analysis, the adapted ESC 0/2 h algorithm had greatest sensitivity. 'Ruling out' AMI using the 99th percentile of the assay cannot be recommended.