Comparison of Javal-Schiøtz keratometer, Orbscan IIz and Pentacam topographers in evaluating anterior corneal topography.
Jennifer M TurnerChristine PurslowPaul J MurphyPublished in: Clinical & experimental optometry (2022)
Algorithms used by Orbscan and Pentacam to solve the peripheral paraxial ray problem produce significantly different KAs. Instrument-specific KAs cannot be used inter-changeably between instruments. Differences in KA between instruments are not significant for ocular surgery, but may influence rigid contact lens fitting. Pentacam measures flatter than Orbscan and J-S.