Estimation of fluid status using three multifrequency bioimpedance methods in hemodialysis patients.
Lin-Chun WangJochen G RaimannXia TaoPriscila PreciadoOhnmar ThwinLaura RosalesStephan ThijssenPeter KotankoFansan ZhuPublished in: Hemodialysis international. International Symposium on Home Hemodialysis (2022)
Although segmental eight-point bioimpedance techniques provided comparable TBW measurements not affected by standing over a period of 10-15 min, the ECW/TBW ratio appeared to be significantly lower in InBody compared with Seca and Hydra. Results from our study showed lack of agreement between different bioimpedance devices; direct comparison of ECW, ICW, and ECW/TBW between different devices should be avoided and clinicians should use the same device to track the fluid status in their HD population in a longitudinal direction.