Anti-Nuclear Antibody Quantitation: Calibration and Harmonization Adjustment via Population Interrogation.

Mark H WenerSusan L FinkChihiro MorishimaAnu ChaudharyKathleen Hutchinson
Published in: The journal of applied laboratory medicine (2022)
One approach to achieving harmonization of ANA assays from different laboratories with each other and with expected performance would involve adjusting assays so that about 10% of a healthy control population has a positive ANA when tested at 1:80 dilution, and about 20% of the healthy control population has a positive ANA when tested at 1:40 dilution. This pragmatic approach to calibration and harmonization adjustment via population interrogation offers an opportunity for individual laboratories to be aligned with each other and with ANA performance expected for consistent categorization of patients with SLE.