Calibration of coherence imaging spectroscopy using spectral line sources.

Kenji UedaMasaki NishiuraNaoki KenmochiZensho YoshidaKaori Nakamura
Published in: The Review of scientific instruments (2021)
Coherence imaging spectroscopy (CIS) measures the two-dimensional profiles of both ion temperature and ion velocity in plasmas. The interferometric technique is realized by a certain relation between the phase and the wavelength of light emerging from a birefringent crystal. The calibration for the CIS system requires monochromatic and tunable light sources near the He II line (468.6 nm) or C III line (465 nm) where the CIS measures. In this research, the CIS system has been upgraded by implementing an electron multiplier CCD and a CIS cell. A monochromator validates the linearity of the phase relation on the wavelength near the He II line. As an in situ calibration at the Ring Trap 1 plasma device, two spectral lines of Ti and Zn lamps obtain the accurate dispersion function of phase. It is found that a simple method with two spectral lines is reliable and sufficient for the calibration.