Login / Signup

Effect of Sinotubular Junction Size on TAVR Leaflet Thrombosis: A Fluid-structure Interaction Analysis.

David OksSymon RezaMariano VázquezGuillaume HouzeauxBrandon KovarovicCristóbal SamaniegoDanny Bluestein
Published in: medRxiv : the preprint server for health sciences (2023)
A smaller STJ size impairs adequate expansion of the TAVR stent, which may lead to suboptimal hemodynamic performance. Conversely, a larger STJ size marginally enhances the hemodynamic performance but increases the risk of TAVR leaflet thrombosis. Such analysis can aid pre- procedural planning and minimize the risk of TAVR leaflet thrombosis.
Keyphrases
  • aortic valve
  • transcatheter aortic valve replacement
  • aortic stenosis
  • mitral valve
  • pulmonary embolism
  • left ventricular
  • coronary artery disease