Complex confounder-corrected R2* mapping for liver iron quantification with MRI.
Diego HernandoRachel J CookNaila QaziColin A LonghurstCarol A DiamondScott B ReederPublished in: European radiology (2020)
• Confounder-corrected R2* of the liver provides reproducible R2* across acquisition protocols, including different spatial resolutions, echo times, and slice orientations, at both 1.5 T and 3.0 T. • For all acquisition protocols, high correlation with R2-based liver iron concentration (LIC) quantification was observed. • The calibration between confounder-corrected R2* and LIC, at both 1.5 T and 3.0 T, is determined in this study.