Login / Signup

Biomechanical Comparison of Use of Two Screws versus Three Screws Per Fragment with Locking Plate Constructs under Cyclic Loading in Compression in a Fracture Gap Model.

Sophie PalierneMargaux BlondelPascal SwiderAndré Autefage
Published in: Veterinary and comparative orthopaedics and traumatology : V.C.O.T (2022)
 Omission of the third innermost locking screw during bridging osteosynthesis subjected to compression forces led to a 13.3% reduction in the number of cycles to failure and a 23.5% reduction of the load withstood by the plate before plastic deformation occurred.
Keyphrases
  • finite element analysis
  • hip fracture
  • clinical evaluation
  • finite element