Login / Signup

First Assessment of a Carbon Monoxide Laser and a Thulium Fiber Laser for Fractional Ablation of Skin.

Linh Ha-WisselMartin JaspanDavid WelfordMichael EversGaruna KositratnaMalte Johannes CasperDieter MansteinReginald Birngruber
Published in: Lasers in surgery and medicine (2020)
The tissue ablation using the Tm:fiber laser was very superficial in contrast to the CO laser and the CO2 laser. However, higher etch depths should be obtainable when the radiant exposure is increased by using higher pulse energies and/or smaller spot sizes. At radiant exposures normalized to the ablation threshold, the CO laser was capable of generating deeper ablation craters with extended coagulation zones compared with the CO2 laser, which is possibly desirable depending on the clinical goal. The effect of deep ablation combined with additional thermal damage on dermal remodeling needs to be further confirmed with in vivo studies. Lasers Surg. Med. © 2020 The Authors. Lasers in Surgery and Medicine Published by Wiley Periodicals, Inc.
Keyphrases