Validation of the accuracy of contact force measurement by contemporary force-sensing ablation catheters.
Thomas KuefferAndreas HaeberlinSven KnechtSamuel H BaldingerAntonio MadaffariJens SeilerAline MühlHildegard TannerLaurent RotenTobias ReichlinPublished in: Journal of cardiovascular electrophysiology (2022)
Contact measured by force-sensing catheters is accurate with 1-3g deviation within the range of 10g to 40g. Significant errors can occur at higher forces with potential clinical consequences. This article is protected by copyright. All rights reserved.